Testa att söka på nyckelord, orderkod eller produktkod, eller serienummer t.ex. ”CM442” eller ”teknisk information”
Enter at least 2 characters to start the search.
Indumax CLS54D is an inductive conductivity probe for the highest hygienic and sterile demands.

Digital conductivity sensor
Indumax CLS54D

Memosens toroidal conductivity sensor for hygienic and sterile applications

CLS54D
Please Wait (spinning wheel)
Laddar pris
Pris ej tillgängligt för tillfället
Price on request
Aktiverar pris
pris: ange information
from CHF ??.-
Filtrera mät-/applikationsuppgift
Mät-/applikationsuppgift

Konfigurera för att ladda ner denna produkts CAD-ritningar

  • Mätprincip

    Inductive

  • Application

    Food, beverage and pharmaceutical industries, biotechnology, dairies, breweries

  • Characteristic

    Specifically intended for use in hygienic applications and especially suitable for following applications:
    Phase separation, control of CIP, concentration control, product monitoring, leakage monitoring

  • Measurement range

    100 µS/cm to 2000 mS/cm
    Cell constant k: 6.3 1/cm

  • Measuring principle

    Inductive conductivity measurement

  • Design

    Hygienic design with food-grade virgin PEEK body in jointless and creviceless design including all hygienic certifications

  • Material

    In contact with medium: virgin PEEK

  • Process temperature

    -10 °C to 125 °C (14 °F to 257 °F)
    Sterilization: 150 °C at 6 bar abs for max. 60 min
    (302 °F at 87 psi abs for max. 60 min)

  • Process pressure

    13 bar abs up to 90 °C (188.5 psi abs up to 194 °F)
    9 bar at 125 °C (130.5 psi abs at 257 °F)

  • Temperature sensor

    Pt1000 (Class A according to IEC 60751)

  • Connection

    Dairy pipe, Aseptic-fitting, Clamp ISO 2852, SMS-2", Varivent N DN 40-125, Neumo Biocontrol D50

  • Ingress protection

    IP68 / NEMA type 6P

  • Additional certifications

    FDA, EHEDG, 3-A
    Biological reactivity USP class VI part <87> and <88>