Testa att söka på nyckelord, orderkod eller produktkod, eller serienummer t.ex. ”CM442” eller ”teknisk information”
Enter at least 2 characters to start the search.
Product picture of resistance thermometer TR13

RTD Thermometer TR13

TR13
Please Wait (spinning wheel)
Laddar pris
Pris ej tillgängligt för tillfället
Price on request
Aktiverar pris
pris: ange information
from CHF ??.-
Filtrera mät-/applikationsuppgift
Mät-/applikationsuppgift

Konfigurera för att ladda ner denna produkts CAD-ritningar

  • Mätprincip

    Resistance Temperature Detector

  • Characteristic / Application

    metric style
    modular temperature assembly
    universal range of application
    suitable for hazardous areas
    flanged process connection
    with neck
    incl. thermowell / protection tube (metal)
    usable with insert StrongSens

  • Thermowell / protection tube

    welded protection tube

  • Insert / probe

    mineral insulated (MI), flexible

  • Outer diameter protection tube / Insert

    9,0 mm (0,35'')
    11,0 mm (0,43'')
    12,0 mm (0,47'')
    14,0 mm (0,55")
    15,0 mm (0,59")

  • Max. immersion length on request

    up to 10.000,00 mm (393,70'')

  • Material protection tube/ thermowell

    1.4404 (316L)
    1.4571 (316Ti)
    Alloy C276 (2.4819)
    Alloy 600 (2.4816)

  • Optional coating

    Tantal
    PTFE Teflon
    PVDF

  • Process connection

    flange:
    DN15 PN40 B1 (EN1092)
    DN15 PN40 C (EN1092)
    DN25 PN20 B1 (EN1092)
    DN25 PN40 C (EN1092)
    DN25 PN50 B1 (EN1092)
    DN25 PN100 B2 (EN1092)
    DN40 PN40 B1(EN1092)
    DN50 PN40 B1 (EN1092)
    ASME 1" 150 RF (B16.5)
    ASME 1" 300 RF (B16.5)

  • Tip shape

    straight
    reduced
    tapered

  • Surface roughness Ra

    1,6 μm (63,0 μin.)

  • Operating temperature range

    PT100 TF StrongSens:
    -50 °C ...500 °C
    (-58 °F ...932 °F)
    PT100 WW:
    -200 °C ...600 °C
    (-328 °F ...1.112 °F)
    PT100 TF:
    -50 °C ...400 °C
    (-58 °F ...752 °F)

  • Max. process pressure (static)

    at 20 °C: 50 bar (725 psi)

  • Accuracy

    class A acc. to IEC 60751
    class AA acc. to IEC 60751

  • Response time

    depending on configuration

  • Integration head transmitter

    yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

  • Ex - approvals

    ATEX II1D Ex ia IIIC, II1G Ex ia IIC T6
    ATEX II1/2D Ex ia IIIC,II1G Ex ia IIC T6
    ATEX IECEx II1/2D Ex ta/tb IIIC Da/Db
    ATEX II1G Ex ia IIC T6
    ATEX II 3 G Ex nA IIC T6, II3D
    NEPSI Ex ia IIC T6, Ex iaD 20 T85-T450
    IECEx Ex ia IIC T6 Ga/Gb
    EAC Ex ia IIC T6 Ga + DIP
    NEPSI Ex nA IIC T6

  • Certification

    Gost Metrology
    SIL (transmitter only)